How we can help

Whether hospitals are looking to support this crucial service during periods of refurbishment, to temporarily increase clinical capacity during peaks in demand , our new mobile endoscopy dual procedure suite could be the innovation that helps to answer the challenge. They offer HEPA-filtered environmental air and include inbuilt decontamination capability with a one-way endoscope flow.

With 2 procedure rooms working in tandem supported by 5 patient bays, and decontamination facilities for up to 9 flexible endoscopes per hour. Providing a spacious, climate-controlled work environment with natural light, designed and equipped in consultation with frontline staff. It can be installed in a matter of hours and becomes operational following a commissioning and validation period.

Q-bital mobile endoscopy dual procedure suite includes:

  • 2 procedure rooms
  • 5 bay recovery ward
  • Flexible endoscope reprocessing rooms
  • Dirty utility area
  • Pass-through thermally sanitised AERs
  • Thermally sanitised Reverse Osmosis plant
  • BS EN 16442 compliant endoscope drying cabinet
  • IP65-rated track and trace system
  • HEPA-filtered environmental air
  • Technical room
  • Admission and discharge areas
  • Scrub area
  • Reception
  • Waiting room
  • Consulting room
  • WC with wheelchair access
  • Staff welfare area
  • Changing room
  • Refreshment area
  • Utility areas

Considerations

Unit dimensions 16m length x 16m width x 3.3m height
Electrical supply 2 x 125 Amp, 3PN+E power supply
Water supply 2 x 22mm and 1 x 15mm tap connections at 3 bar pressure
Foul drainage 5 x 50mm flexible pipes and 1 x 45mm pipe
Telecommunications NTE5 (British Telecoms standard) telephone connection. Data cable network.
Other

Fire alarm cable zero-volt connections

On-board oxygen cylinder manifolds

On-board vacuum plant

Ward area has IPS electric sockets

1 x staff WC

1 x patient toilet with wheelchair access

The installation process takes 3-4 hours, followed by an agreedĀ commissioning period.